Cleaning of submersible pressure transmitters or level probes

If the specific pressure sensor design of the submersible pressure transmitter or level probe is selected to measure the filling levels, this often means that the probe is used under environmental requirements which would cause failure of common level sensors.
The most unfortunate circumstances such as soiled media, abrasive ingredients and sludge when used in wastewater treatment plants, brackish and wastewater tanks and even digester towers, impose special requirements on the look of a submersible pressure transmitter. One of many requirements on a submersible pressure transmitter is to have the lowest possible susceptibility to contamination or build-up of the pressure sensor by optimizing its design. For this reason the typical design of a pressure transmitter with narrow pressure ports isn’t used within level probes because it would tend to clog in such applications.
ไดอะแฟรม of the submersible pressure transmitter and its own pressurised sensor diaphragm is optimised in order to achieve suprisingly low susceptibility to contamination. However, continuous operation in soiled media may lead to sticking of dirt particles on the stainless diaphragm. To get the highest accuracy and fastest response times in case of level change, the thickness of the stainless steel diaphragm has already been minimised ex factory to just a couple microns. Therefore, cleaning of the diaphragm must be completed with caution. Always stay away from sharp or edged tools. Additionally it is strongly advised not to use the commonly used screwdrivers or pens.
If cleaning of the sensor diaphragm is essential, then rinse it utilizing a weak water jet or clean it carefully using compressed air. Damage of the diaphragm because of denting or notching, even though it appears to be purely superficial, results in significant losses in the accuracy of level measurement. Deformation of the diaphragm often shifts the zero point of the pressure measurement in the internal electronic measurement system and additionally distorts the output signal linearisation which has been adjusted ex works to the undamaged diaphragm. Thus, the submersible pressure transmitter with damaged diaphragm generates falsified measurement of the existing filling level and, therefore, can’t be considered a reliable measuring instrument any longer. Thus, complete replacement of the damaged instrument is absolutely necessary.
Please find more info on this topic on our information platform ?Hydrostatic level measurement?

Scroll to Top